Remove the cartridge protective cap.
Make sure that you do not touch the cartridge tip.
Push the adapter onto the cartridge tip as far as to the
stop (a).
Carefully prepare a new Accu-Chek infusion set
for use.
Hold the adapter and twist the infusion set by hand
clockwise into the adapter (b).
Do not overtighten.
a
b
Luer-lock connector
Warning
The cartridge and infusion set are
sterile products
intended for single use only.
Sterility is guaranteed for
unopened packaging up to the indicated expiration date.
Do not use sterile products if the packaging is
damaged.
Do not reuse single use materials. Reuse of single use
materials might lead to a malfunction of your insulin
pump and to incorrect insulin delivery and/or an
infection. Always handle the items with clean hands.
Avoid any contact of the infusion set and especially the
connecting parts of your insulin pump with antiseptics,
antibiotic creams, soaps, perfumes, deodorants, body
lotions or any other cosmetics. They may contaminate
these parts.
59
Prepare the pump for use