10.2.11 Error E12: DATA INTERRUPTED
When data transfer between your insulin pump and a PC is interrupted an
error E12: DATA INTERRUPTED occurs. Restart communication between
your insulin pump and the PC and make sure that the data transfer is
completed correctly. Before you can put your insulin pump into
RUN
,
the data transfer must be successfully concluded.
Refer to the Accu-Chek Insulin Pump Configuration Software user manual
for further instructions on using the data transfer feature.
Press
f
twice to turn off and confirm the error.
Restart the data transfer.
Caution
If data transfer between your PC and your insulin pump
is disrupted in any way, the configuration may be
incomplete and an error E12: DATA INTERRUPTED
occurs. The data transfer must be completed success-
fully before you can put your insulin pump into
RUN
.
163