7.8.2 Review the alarm history
The alarm history enables you to review the last 30 alerts or errors
starting with the most recent entry in chronological order.
Each alarm history screen displays the
• number and type of the alert or error (e.g.
A6),
• alert or error (e.g. TBR CANCELLED),
• time (
),
• date (
) and
• entry number (01 is the most recent) of total entries (e.g. 01/30)
of an alert or error.
Press
d
to move to the INFORMATION menu.
Select with
f
.
Press
d
to move to the ALARM HISTORY
screen.
Press
a
or
s
to review the entries.
Exit with
f
.
7.8.3 Review the daily insulin totals history
This history enables you to review the last 30 daily totals of insulin
delivered (from midnight to midnight; basal rate plus bolus deliveries)
starting with the most recent entry in chronological order.
Each daily insulin total screen displays the following history
• daily total of insulin delivered (
),
• date (
) and
• entry number (01 is the most recent) of total entries (e.g. 01/30)
of a daily insulin total.