Note 1
At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2
These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from
structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for
radio (cellular/cordless) telephones and land mobile radios, ama-
teur radio, AM an FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To asses the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the
location in which the Accu-Chek Spirit is used exceeds the applica-
ble RF compliance level above, the Accu-Chek Spirit should be
observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorient-
ing or relocating the Accu-Chek Spirit.
b
Over the frequency range 150 kHz to 80 MHz, field strengths
should be less than 10 V/m.
Rated maximum
output power
of transmitter W
80 MHz to
800 MHz
800 MHz to
2.5 GHz
V
1
in V:
10.00
E
1
in V/m:
10.00
150 kHz to 80 MHz
out of ISM
(Industrial, Scientific,
Medical Band)
Separation distance according to frequency of transmitter m
0.04
0.11
0.35
1.11
3.50
0.07
0.22
0.70
2.21
7.00
0.04
0.11
0.35
1.11
3.50
0.01
0.1
1
10
100
193
Technical data
Summary of Contents for Spirit
Page 1: ...INSULIN PUMP User Guide ...
Page 10: ...8 ...
Page 14: ...12 ...
Page 26: ...24 ...
Page 36: ...34 ...
Page 76: ...74 ...
Page 84: ...82 ...
Page 122: ...120 ...
Page 146: ...144 ...
Page 186: ...184 ...
Page 212: ...210 ...
Page 218: ...216 ...
Page 225: ......