Press
f
twice to turn off and confirm the alert.
Put your insulin pump into
RUN
.
Check the INFORMATION screen on your pump
for the PUMP TIMER display. The amount shown
is the remaining number of programmed days in
your insulin pump.
After the alert A5: PUMP TIMER occurs, the
Accu-Chek Spirit insulin pump will display the
symbol to remind you to check the timer.
Contact your local HelpDesk to discuss your
options for further use of the Accu-Chek Spirit
insulin pump.
Note
The alert A5: PUMP TIMER and the error E5: END OF
OPERATION are based on running time (period with pump in
RUN
) and not calendar years. Your actual warranty is based
upon the date the pump was first purchased.
10.1.6 Alert A6: TBR CANCELLED (Temporary Basal Rate cancelled)
An alert A6: TBR CANCELLED occurs when a temporary increase or
decrease of the basal rate has been cancelled. The original basal rate
(100 %) will resume automatically.
Press
f
twice to turn off and confirm the alert.
Put your insulin pump into
RUN
, if necessary.
Ensure that the cancellation was intended and, if
necessary, program a new temporary basal rate.