7.5.4 Program a menu-guided “Scroll” Standard bolus
The bolus increments for the “Scroll” Standard bolus are fixed to 0.1 unit.
You can program this bolus in the STANDARD BOLUS menu by holding
the
a
or
s
key (scrolling) until the desired bolus amount is displayed.
Press
d
to move to the STANDARD BOLUS
menu.
Select with
f
. The BOLUS AMOUNT screen
appears.
Press or scroll
a
to increase or
s
to
decrease the bolus amount.
Press
f
to confirm the bolus amount.
During 5 seconds
h
(bolus delivery start delay),
the Standard bolus symbol (
) blinks.
After this bolus delivery start delay, your insulin
pump beeps three times and starts to deliver the
total bolus amount programmed. The countdown
of the remaining bolus amount appears on the
display.