Alerts and errors
10.2.2 Error E2: BATTERY DEPLETED
If the battery is depleted, an error E2: BATTERY DEPLETED occurs.
Replace it immediately.
Press
f
twice to turn off and confirm the error.
Replace the battery.
When finished, put your insulin pump into
RUN
.
Note
Your insulin pump stores the settings you made and saved
before the battery was depleted. If your insulin pump
remains without power for more than one hour, check the
time and date.
10.2.3 Error E3: AUTOMATIC OFF
This safety function interrupts insulin delivery if no keys are pressed
within a certain period of time.
Refer to section 7.7.4 “Automatic off” for more information.
Press
f
twice to turn off and confirm the error.
Put your insulin pump into
RUN
.
Warning
After inserting a new battery, always ensure that the
time and date of your insulin pump are set correctly.
Incorrect programming of the time and date may cause
incorrect insulin delivery. If you or your healthcare team
store and analyse your therapy data electronically and
the time and date of the devices used are not set
identically, the gathered data might not be meaningful.