Press
d
to move to the ALARM SIGNALS
screen. The active alarm signals are displayed.
Press
a
or
s
to set the alarm signals as
desired.
Save and exit with
f
.
7.7.4 Automatic off
Automatic off is a safety feature which stops all insulin delivery by
triggering an error E3: AUTOMATIC OFF if no keys are pressed within a
programmed time period in
RUN
.
The automatic off can either be set to OFF or programmed up to 24 hours
in 1 hour intervals in the SETUP MENU STANDARD.
Press
d
to move to the SETUP MENU
STANDARD.
Select with
f
.
Press
d
to move to the AUTOMATIC OFF
screen. The active status is displayed.
Press or scroll
a
to increase or
s
to
decrease the duration for the automatic off in 1
hour intervals. If you would like to set the auto-
matic off function to OFF, press or scroll
s
until 0 hours and OFF appear on the display.
Save and exit with
f
.
Caution
Discuss the use of the automatic off function with your
doctor or healthcare team.
111
STANDARD user menu
Summary of Contents for Spirit