Remove the battery cover from your insulin
pump. Use the battery key to turn the battery
cover counter-clockwise as indicated by the
arrow on the battery cover.
Make sure the opening of the battery
compartment and the seal are clean and
undamaged.
Put the battery cover on the positive end of the
battery.
Insert the battery with cover, with the flat
(negative) end first, into the battery
compartment.
Carefully push the battery with cover in and turn
the battery cover clockwise. The battery cover is
correctly inserted and tightened when the battery
cover is even with the casing. Do not over-
tighten.
Your insulin pump will now perform its start-up
procedure.
A melody signals the end of the start-up proce-
dure, and your insulin pump returns to the
STOP
screen.