10.2.5 Error E5: END OF OPERATION
The error E5: END OF OPERATION indicates that your pump timer has
reached zero. The pump will no longer operate. Before this error occurs
an alert A5: PUMP TIMER will occur to ensure that you are able to replace
your pump within sufficient time.
When the PUMP TIMER has reached zero, it will
change into
STOP
and will no longer operate. An
E5: END OF OPERATION will appear on the
display.
Press
f
twice to turn off and confirm the error.
To ensure that insulin delivery is optimized, your
pump features a pump timer.
To inspect the pre-set lifetime of your Accu-Chek
Spirit pump before you start using it, please refer
to the “Accu-Chek Spirit Pump Protocol”
included in the starter kit. During use, the
remaining lifetime in days until the expiration
date of your insulin pump can be reviewed
in the INFORMATION menu (see “Review the
time remaining”).
Contact your local HelpDesk to discuss your
options for further use of the Accu-Chek Spirit
insulin pump.
Note
The alert A5: PUMP TIMER and the error E5: END OF
OPERATION are based on running time (period with pump in
RUN
) and not calendar years. Your actual warranty is based
upon the date the pump was first purchased.