To confirm an alert or error
Press
f
to turn off the beeps and vibrations.
The alert or error code remains on the display.
Confirm that you have seen and understood the
alert or error code and messages by pressing
f
again.
Take the appropriate action, if necessary.
If more than one alert and/or error occur at the same time, press
f
twice for each alert or error to turn off and confirm.
When you confirm A1: CARTRIDGE LOW
, E1: CARTRIDGE
EMPTY
, A2: BATTERY LOW
, E2: BATTERY DEPLETED
,
A5: PUMP TIMER
or E5: END OF OPERATION, the symbol for
that alert and error stays in the
RUN
screen or
STOP
screen as a
reminder. For the other alerts and errors, the alarm information
is cleared from the display and remains only in the alarm history
and the event memory.
After the occurrence of an alert or error, it is important to make sure that
your insulin pump is returned to
RUN
, if desired.