197
Technical data
Basal rate
– hourly basal rate + max.
TBR (hourly basal rate
combined with temporary
basal rate, max. increase)
– hourly basal rates
increment
– hourly basal rates
– temporary basal rate
TBR increase or decrease
in increments of
TBR duration
TBR duration increments
– basal rate profiles available
– basal rate lock
Bolus
– max. bolus amount per
bolus delivery
– bolus amount increments
for “Quick” Standard bolus
– bolus duration (Extended
bolus, MultiWave bolus)
in increments of
Prime volume
Typical range available
on the pump and
standard parameter
settings*
20.0 IU/h
0.1 IU
10.0 IU/h
0 – 200 %
10 %
15 min – 24 h
15 min
1 – 5
on or off
presetting: off
25.0 IU
0.1, 0.2, 0.5, 1.0 or 2.0 IU
presetting: 0.5 IU
15 min – 12 h
15 min
0 – 30.0 IU
presetting: 25 IU
Maximum parameter range
adjustable with the Accu-Chek
Insulin Pump Configuration
Software
max. amount
0 – 62.50 IU/h
max. amount
0 – 25.0 IU/h
0 – 250 %
10 %
15 min – 24 h
15 min, 30 min, 1 h
1 – 5
on or off
max. amount
0 – 25.0 IU
0.1, 0.2, 0.5, 1.0 or 2.0 IU
15 min – 24 h
15 min, 30 min, 1 h
max. amount
0 – 30.0 IU
* Note: these factory settings may vary in some countries and depend on the changes you,
your doctor or healthcare team made using your insulin pump or pump programming
software for healthcare professionals.
Summary of Contents for Spirit
Page 1: ...INSULIN PUMP User Guide ...
Page 10: ...8 ...
Page 14: ...12 ...
Page 26: ...24 ...
Page 36: ...34 ...
Page 76: ...74 ...
Page 84: ...82 ...
Page 122: ...120 ...
Page 146: ...144 ...
Page 186: ...184 ...
Page 212: ...210 ...
Page 218: ...216 ...
Page 225: ......