HR-501 

Heart Guardian AED User Manual 

HR-501 operation Manual Rev.1.1                                                                      Rev. 2016/02/12 

 

Summary of Contents for Heart Guardian AED

Page 1: ...HR 501 Heart Guardian AED User Manual HR 501 operation Manual Rev 1 1 Rev 2016 02 12...

Page 2: ...ate product operation Contents of the manual may be changed without prior notification or agreement by Radian Data without any quotation mark may be different from estimated data as compared to actual...

Page 3: ...2 Getting started 4 3 Preparation 4 3 1 AED operation 4 3 2 Data management 4 3 3 Guidance for maintenance Chapter 5 AED location and periodical maintenance 5 1 After first time use 5 2 Troubleshooti...

Page 4: ...ards that may cause an electric shock product damage and malfunction Reference Additional reference data is referred Dangerous Voltage The defibrillator output has high voltage and may present a shock...

Page 5: ...pose to direct sunlight WEEE Waste Electronic Electrical Equipment Separate collection for waste electrical and electronic equipment Direct Current DC Direct Current Battery Type Lithium manganese oxi...

Page 6: ...ctive 93 42 EEC Date of Manufacture 4 digits of the year and 2 digits of month and day which indicate the manufacturing date Manufacture Name and address of manufacturer Bluetooth AED supports Bluetoo...

Page 7: ...ic shock to patient In child mode about 50J of electric shock is set to be delivered if patient is under the age of 7 or weighting below 25kg please set Child Mode by moving Adult child mode switch to...

Page 8: ...e contact local national sales representative for repair or replacement Do not use battery not provided by RADIAN RADIAN is not responsible for damage malfunction in operation caused by unofficial bat...

Page 9: ...our of storage at 20 30 temperature condition prior to actual AED operation For long term storage over 5 days RADIAN recommends maintaining AED with temperature range from 20 to 30 AED pads and batter...

Page 10: ...R 501 operation on a simulator or a mannequin may cause serious malfunction or damage on AED Compliance with national and regional regulation AED maintenance and authorization responsibility of usage...

Page 11: ...roper operation not following the user manual 3 2 2 Defect caused by unauthorized repair or modification 3 2 3 Defect caused by external shock including drop 3 2 4 Defect caused by natural disaster su...

Page 12: ...he defect Once Radian confirm the defect is caused by one of the manufacturing development or delivery procedure repair exchange may be provided without any shipping charge Shipping will be completed...

Page 13: ...g training schedule 4 1 Product overview HR 501 a portable electric device operated by battery power is an easy to use Semi Automated External Defibrillator The AED automatically analysis patient s EC...

Page 14: ...Provides electrics to AED G AED Pad Connector Connect the AED pads prior to attaching the pads to patient H Shock Button Press the button when guidance indicates to deliver charged electric shock to...

Page 15: ...put Range 150J 10 Adult 50J 10 Child Data Storage 108 times of ECG log can be recorded Control Button Power Analysis Shock Adult child mode switch Battery Type BT 303 LiMnO2 12V DC 4 2A Number of shoc...

Page 16: ...ve pads for both adult child Size 145 H x 98 W x 1 T Weight 120g Warranty 2 years in case of damage caused by user s mistake repair exchange will be charged Life time Single use only 2 years for unpac...

Page 17: ...AED operating training schedule Heart Guardian HR 501 is an emergency medical device used for patient stops in breathing lapsed into unconsciousness due to sudden cardiac arrest No response Unusual or...

Page 18: ...ational local representative Manual and quick guide is available at www radian co kr for downloading I HR 501 AED II HR 501 User Manual III P 303 AED Pads IV BT 303 Battery Non rechargeable V Quick Gu...

Page 19: ...nd child If it is hard to define the patient s weight and age quickly proceed further operation procedure with the default mode adult mode Adult Mode Above age of 8 or Minimum weight 25kg Child Mode B...

Page 20: ...ads out 2 Remove the covering package and tear off the covering stickers 3 Attach the pads to patient s chest 4 Connect the pads to AED 4 3 2 5 ECG Analysis Once AED pads are clearly attached to the p...

Page 21: ...ge shock button Electric shock delivers to the patient After electric shock perform CPR procedure following audio guidance Audio guidance Start CPR Audio guidance Compress the chest 30 times Beeper ri...

Page 22: ...roduct operates as below Audio guidance Firmly attach the pads to the patient s bare skin Even though electric energy charge is completed if user does not push blinking shock button for more than 15 s...

Page 23: ...where AED can be easily detected and accessed RADIAN strongly suggests setting AED close to a telephone in order to call emergency aid service immediately in case of emergency 5 1 2 All components in...

Page 24: ...mental weather condition 5 1 6 Battery and AED pads maintenance Battery maintenance If the battery is low AED may not be operated properly in emergency situation When status LED blinks RED battery pac...

Page 25: ...ible for damage malfunction in product operation caused by use of unofficial battery Since use of unofficial battery may cause trouble malfunction AED pads maintenance AED pads expiration date shall b...

Page 26: ...he battery to new one Suggested battery life is 4 years for standby mode 6 Troubleshooting 6 1 Self diagnosis Power On Self test Operation mode Every time when power button is pressed Flash ROM test C...

Page 27: ...link 3 beep sounds Min Prompts when error is found during operation Status Indication Light Green Blink Prompts AED to conduct self diagnosis test Status Indication Light Off Maintenance manager mode...

Page 28: ...n Replace the battery to new one Audio message The battery is low please change the battery immediately Replace the battery to new one Audio message System error 000X Contact the vender for A S servic...

Page 29: ...ue to disconnection of the pads Connect AED pads connector to AED firmly and properly Audio message Plug the connector of the pad to AED repeat Do not attempt to disassemble or repair the AED arbitrar...

Page 30: ...rm storage over 5 days Operation 0 40 32 104 20 30 68 86 0 40 32 104 Humidity 5 95 No saturated place Altitude 0 to 4 575m 0 to 15 000ft AED with pads stored over 5 days at temperatures condition betw...

Page 31: ...er Operating time For a new fully charged battery pack at 20 20 C 200 discharges at maximum energy or 10 hours of continuous ECG monitoring Automated Self tests Monthly every 1st day for each months B...

Page 32: ...Hz Free fall drop IEC60068 2 32 1m Enclosure Particle and water ingress IEC 60529 IP54 EMC IEC60601 1 2CEMC EMC Sensitivity and Specificity of Rhythm Detection FFT ECG Analysis Shockable Rhythm VF IEC...

Page 33: ...tional regulations Shelf life 4 years from date of manufacture can be reduced by temperature condition 7 7 FFT ECG Analysis FFT ECG Analysis algorithm is for efficient and accurate ECG detection algor...

Page 34: ......

Page 35: ......

Page 36: ...1610 Halla Sigma Valley 53 Gasan Digital 2 ro Geumcheon gu Seoul Korea 82 2 1670 1980 Info radian co kr www radian co kr...