Maximum volume before error
E4: OCCLUSION**
Maximum pressure
Flow (delivery rate)
Maximum quantity delivered at
a single fault condition
Cartridge
Infusion sets
Data transfer
Plastic cartridges:
3.5 U
The typical volume for an error
was 2.3 U.
Aventis Insuman Infusat
glass cartridges*:
6.0 U
400 kPa (4.0 bar),
for plastic and Aventis Insuman Infusat
glass cartridges*.
During infusion set priming and bolus
0.2 U/sec.
1.0 U
Accu-Chek 3.15 ml plastic cartridges
with a luer connection and by Roche
Diagnostics approved third party
products. Aventis Insuman Infusat
cartridges* have been evaluated and
approved for use with the Accu-Chek
Spirit insulin pump.
Accu-Chek infusion sets with a luer-lock
connector.
Accu-Chek Infusion sets are PCV and
latex free.
Infrared interface
Technical data
*
*
Insuman
®
Infusat produced by Aventis Pharma Deutschland GmbH, a member and a brand
of the sanofi-aventis group.
** Determined using measurement methods according to IEC 60601-2-24-1998.