8.5.5 Lock basal rate profiles
Your insulin pump allows you to change your hourly basal rates as
needed. This feature provides additional protection against accidentally
changing a basal rate during normal operation. When this function is
activated, your hourly basal rates cannot be changed.
Make sure the SETUP MENU ADVANCED is
turned on.
Press
d
to move to the SETUP MENU
ADVANCED.
Select with
f
.
Press
d
to move to the BASAL RATE LOCK
screen.
Press
a
or
s
to turn the basal rate
lock ON ( ) or OFF ( ).
Save and exit with
f
.
Note
If the basal rate lock is set to ON ( ), the programing of the
basal rate profiles 1, 2, 3, 4 and 5 is blocked. To check if the
basal rate lock is ON, move with
d
to a PROGRAM BASAL
RATE menu and select with
f
. The
symbol in the BASAL
RATE TOTAL screen indicates the lock of the basal rate pro-
graming (The
symbol appears in the BASAL RATE TOTAL
screen only). Using
d
the hourly basal rates can be viewed
but they can not be changed by using
a
and
s
.
138
ADVANCED user menu