15.3 Technical standards on electromagnetic immunity
IEC 60601 test
level
± 6 kV contact
± 8 kV air
3 A/m 400 A/m
Compliance level
± 8 kV contact
± 15 kV air
(IEC 60601-2-24)
400 A/m
(IEC 60601-2-24)
Electromagnetic
environment –
guidance
Avoid any contact
with synthetic
materials. A relative
humidity of 10 %
will allow ESD to be
more likely.
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.
Guidance and manufacturer’s declaration – electromagnetic immunity
The Accu-Chek Spirit is intended for use in the electromagnetic environment
specified below. The customer or the user of the Accu-Chek Spirit should assure
that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
Guidance and manufacturer’s declaration – electromagnetic immunity
The Accu-Chek Spirit is intended for use in the electromagnetic environment
specified below. The customer or the user of the Accu-Chek Spirit should assure
that it is used in such an environment.
Immunity test
Conducted RF
IEC 61000-4-6
IEC 60601
test level
3 Vrms
150 kHz to
80 MHz
Compliance
level
Not applicable
Electromagnetic environment –
guidance
191
Technical data
Summary of Contents for Spirit
Page 1: ...INSULIN PUMP User Guide ...
Page 10: ...8 ...
Page 14: ...12 ...
Page 26: ...24 ...
Page 36: ...34 ...
Page 76: ...74 ...
Page 84: ...82 ...
Page 122: ...120 ...
Page 146: ...144 ...
Page 186: ...184 ...
Page 212: ...210 ...
Page 218: ...216 ...
Page 225: ......