19 Alert overview
In all cases, turn off and confirm the alert prior to addressing the issue.
223
No Alert
A1
CARTRIDGE LOW
A2 BATTERY
LOW
A3
REVIEW TIME AND DATE
A4
ALARM CLOCK
A5 PUMP
TIMER
A6 TBR
CANCELLED
A7 TBR
OVER
A8 BOLUS
CANCELLED
What you have to do
Change cartridge before it is
completely empty.
Replace the battery as soon as
possible.
Set the time and date.
Put your Accu-Chek Spirit into
RUN
if
necessary.
In your INFORMATION menu check for
the insulin pump operation remaining
under the menu PUMP TIMER.
Arrange for replacement of pump
within the remaining pump operation
time (60 days).
Put your Accu-Chek Spirit into
RUN
if
necessary. Ensure the cancellation was
intended and program a new
temporary basal rate if necessary.
Decide whether a further temporary
basal rate change is appropriate and
program one if necessary.
Put your Accu-Chek Spirit into
RUN
if necessary. Ensure that the cancella-
tion was intended and program a new
bolus if necessary.
Page
147
148
148
149
149
150
151
151
No Error
E1 CARTRIDGE
EMPTY
E2 BATTERY
DEPLETED
E3 AUTOMATIC
OFF
E4 OCCLUSION
E5
END OF OPERATION
E6 MECHANICAL
ERROR
E7 ELECTRONIC
ERROR
E8 POWER
INTERRUPT
E10 CARTRIDGE
ERROR
E11 SET NOT PRIMED
E12 DATA
INTERRUPTED
E13 LANGUAGE
ERROR
What you have to do
Change the cartridge.
Change the battery.
Put your Accu-Chek Spirit in
RUN
if
necessary.
See the appropriate section in this user
guide for detailed information.
Arrange for immediate replacement of
your insulin pump. Contact your
healthcare professional for alternative
insulin therapy options.
See the appropriate section in this user
guide for detailed information.
See the appropriate section in this user
guide for detailed information.
See the appropriate section in this user
guide for detailed information.
See the appropriate section in this user
guide for detailed information.
Prime your infusion set.
Restart the data transfer.
See the appropriate section in this user
guide for detailed information.
Page
152
153
153
154
157
158
159
160
161
162
163
164
Summary of Contents for Spirit
Page 1: ...INSULIN PUMP User Guide ...
Page 10: ...8 ...
Page 14: ...12 ...
Page 26: ...24 ...
Page 36: ...34 ...
Page 76: ...74 ...
Page 84: ...82 ...
Page 122: ...120 ...
Page 146: ...144 ...
Page 186: ...184 ...
Page 212: ...210 ...
Page 218: ...216 ...
Page 225: ......