• Put your insulin pump into
STOP
.
The temporary basal rate delivery is interrupted.
An alert A6: TBR CANCELLED occurs. Press
f
twice to confirm
and turn off the alert.
See the section 10.1.6 “Alert A6: TBR CANCELLED” for further
information.
Note
If an Extended bolus or a MultiWave bolus is active at the
same time, it is also cancelled when your insulin pump is put
into
STOP
. An alert A8: BOLUS CANCELLED and an alert
A6: TBR CANCELLED occur.
Press
f
twice to confirm and turn off the first alert. Then
the second alert appears on the display. Press
f
twice to
confirm and turn off the second alert. Both alerts are
recorded in the alarm history.
See the sections 10.1.6 “Alert A6: TBR CANCELLED”
and 10.1.8 “Alert A8: BOLUS CANCELLED” for further
information.
Ensure that the cancellation was intended and program a new temporary
basal rate (and/or an Extended or MultiWave bolus), if necessary.
Caution
When your insulin pump is in
STOP
, it does not deliver
any insulin. Put your insulin pump into
RUN
to continue
the insulin delivery.