7.5.1 Standard bolus
Your insulin pump offers you two possibilities to program a Standard
bolus:
“Quick” Standard bolus
“Scroll” Standard bolus
Using the
a
and
s
keys of your
Menu-guided using the
d
and
f
insulin pump, with bolus increments
keys with the
a
and
s
keys for
determined by the user.
programming of the amount.
The bolus amount per delivery is limited to 25 units
h
of insulin.
Programming a Standard bolus includes a short start delay of 5 seconds
before the actual bolus delivery. If necessary, this allows you to cancel the
bolus before the bolus delivery has started by the press of the
a
or
s
key. Alert A8: BOLUS CANCELLED will occur. Press
f
twice to confirm
and turn off the alert. See the section 10.1.8 “Alert A8: BOLUS
CANCELLED” for further information.