ScreenMaster RVG200
Paperless recorder
8 Specification
142
OI/RVG200-EN
Inputs
The figures in the following table include linearizer and electrical errors
Linear inputs
Standard analog input
Accuracy
(% of reading)
Millivolts
Milliamps
Volts
Resistance
(low)
Resistance
(high)
–150 to 150 mV
–50 to 50 mA
–10 to 24 V*
0 to 550
0 to 10000
0.1 % or ±20 µV
0.1 % or ±10 µA
0.1 % or ±10 mV
0.1 % or ±0.5
0.1 % or ±5
Sample interval
125 ms per sample
(all inputs are processed in parallel)
Channel-to-channel input isolation
Galvanically isolated to 500 V DC
Isolation from rest of recorder
Galvanically isolated to 500 V DC
Maximum range
Measurement accuracy
(% of reading)
Thermocouple
ºC
ºF
B
250 to 1800
482 to 3272
0.1 % or ±1 ºC (1.8 ºF)
C
0 to 2300
32 to 4172
0.1 % or ±0.5 ºC (0.9 ºF)
D
0 to 2310
32 to 4190
0.1 % or ±1.5 ºC (2.7 ºF)
E
–100 to 900
–148 to 1652
0.1 % or ±0.3 ºC (0.54 ºF)
J
–100 to 900
–148 to 1652
0.1 % or ±0.3 ºC (0.54 ºF)
K
–100 to 1300
–148 to 2372
0.1 % or ±0.3 ºC (0.54 ºF)
L
–100 to 900
–148 to 1652
0.1 % or ±0.3 ºC (0.54 ºF)
N
–200 to 1300
–328 to 2372
0.1 % or ±0.3 ºC (0.54 ºF)
R
–50 to 1700
–58 to 3092
0.1 % or ±0.3 ºC (0.54 ºF) (above 300 ºC [572 ºF])
S
–50 to 1700
–58 to 3092
0.1 % or ±0.3 ºC (0.54 ºF) (above 200 ºC [392 ºF])
T
–200 to 300
–328 to 572
0.1 % or ±0.3 ºC (0.54 ºF)
RTD
PT100
–200 to 600
–328 to 1112
0.1 % or ±0.5 ºC (0.9 ºF)
PT1000 (IEC 60 751)
–200 to 850
–328 to 1562
0.1 % or ±0.5 ºC (0.9 ºF)
Ni120
–80 to 260
–112 to 500
0.1 % or ±0.5 ºC (0.9 ºF)
Ni1000
–30 to 130
–22 to 266
0.1 % or ±0.5 ºC (0.9 ºF)