ScreenMaster RVG200
Paperless recorder
4 Operation
OI/RVG200-EN
39
Operator Message
Chart, indicator and
bargraph views only
Note.
Greyed-out if the relevant
Operator message
menu enable is not
selected – see pages 83 (Chart view), 84 (Indicator view) or 85 (Bargraph
view) as applicable.
Touch to add one of 24 predefined Operator Messages (see Section 7.7.7,
page 79) or one User-Defined Message to the Alarm Event log.
If
< user defined >
is selected, a data entry keyboard is displayed to enable
the message to be entered (see Fig. 7.6, page 65).
The selected or user-defined message is displayed briefly on screen. If
Operator Messages
annotation is selected (see
Chart Annotation
on
page 82) the message is also added to the chart.
Note.
When the recorder is in Historical Review mode, Operator Messages
generated are added at the current time, not the time indicated by the
cursor.