ScreenMaster RVG200
Paperless recorder
8 Specification
OI/RVG200-EN
141
Archiving to removable media
SD card size
Cards up to 32 GB capacity may be used
USB flash drive size
Drives up to 32 GB capacity may be used
Archive media compatibility
ABB
recorders
comply
with
approved
industry
standards for SD cards and USB flash drives. ABB fully
tests the brands of SD cards and USB flash drives that
it supplies. Other brands may not be fully compatible
with this device and therefore may not function
correctly.
Recording duration
Approximate
duration
calculated
for
continuous
recording of 6 channels of analog data (for example, for
12 channels divide by 2, for 3 channels multiply by 2).
Analog input modules
General
Data that can be
saved to removable
media:
Recorded data per channel
(1 to 24)
Alarm event log data
Totalizer log data
Audit log data
Configuration
File structure:
Binary encoded
File protection:
Secure binary format with data
integrity checks
New file generation
interval:
Automatic
Archive sample
rates:
Data is archived at the same
sample rate at which it is stored
internally
Filename:
20-character tag, prefixed with
date / time
Data verification:
Carried out automatically on all
writes to removable-media files
Duration
Sample rate
512 MB SD Card
1 GB SD Card
1 seconds
8 months
16 months
10 seconds
6 years
13 years
40 seconds
26 years
51 years
60 seconds
40 years
75 years
120 seconds
80 years
255 years
480 seconds
315 years
620 years
Number of process
inputs:
6 per module, maximum of
24 inputs
Input types:
mA, mV, voltage, resistance,
thermocouple, RTD,
digital volt-free, digital 24 V
Thermocouple
types:
B, C, D, E, J, K, L, N, R, S, T
Resistance
thermometer:
PT100, PT1000, Ni120, Ni1000
Other
linearizations:
x, x
3
/
2
, x
5
/
2
, custom
linearization
Digital filter:
Programmable 0 to 60 seconds
Display range:
–999999 to 9999999
Common mode
noise rejection:
>120 dB at 50 / 60 Hz with
300
imbalance resistance
Normal (series)
mode noise
rejection:
>60 dB at 50 / 60 Hz
CJC rejection ratio: ±0.05 ºC / ºC
CJC error 0.5 °C maximum with
recorder @ 25 °C
Sensor break
protection:
Programmable as upscale or
downscale
Temperature
stability:
0.02 % / ºC or 2 µV / ºC
(non-thermocouple ranges only)
AMS 2750 E:
Subject to suitable field
calibration, meets the
requirements of 'Control,
Monitoring and Recording
Instruments' and 'Field Test
Instruments'
Long term drift:
<0.1 % of reading or
20 µV annually
Input impedance:
>10 M
(millivolts inputs)
>900 k
(voltage inputs)
10
(mA inputs)