Note
Action
Follow the instructions given in the refer-
ence calibration routine on the FlexPendant
to create reference values.
If the robot is to be calibrated with refer-
ence calibration:
Find previous reference values for the axis
or create new reference values. These val-
ues are to be used after the repair proced-
ure is completed, for calibration of the ro-
bot.
Creating new values requires possibility to
move the robot.
Read more about reference calibration for
Axis Calibration in
.
If no previous reference values exist, and
no new reference values can be created,
then reference calibration is not possible.
If the robot is to be calibrated with fine
calibration:
Remove all external cable packages
(DressPack) and tools from the robot.
Removing the gearbox
Use these procedures to remove the gearbox.
CAUTION
When performing these procedures, the cable harness will still be fitted or partly
fitted to the robot. Use extreme caution not to cause any damage to the cable
harness.
Lifting down the robot from inverted position
Note
Action
Suitable screws, lightly lubricated:
DANGER
The robot must always be secured to the founda-
tion if any kind of repair or maintenance work is
to be performed.
For some repair work support legs are required.
1
M24 x 100 (min. 4 pcs)
For hole configuration, see
configuration, base on page 74
DressPack can stay fitted.
Remove any payload and tools from the robot.
2
Continues on next page
Product manual - IRB 5720
473
3HAC079195-001 Revision: A
© Copyright 2022 ABB. All rights reserved.
5 Repair
5.9.2 Replacing the axis-2 gearbox
Continued
Summary of Contents for IRB 5720
Page 1: ...ROBOTICS Product manual IRB 5720 ...
Page 8: ...This page is intentionally left blank ...
Page 14: ...This page is intentionally left blank ...
Page 36: ...This page is intentionally left blank ...
Page 186: ...This page is intentionally left blank ...
Page 580: ...This page is intentionally left blank ...
Page 604: ...This page is intentionally left blank ...
Page 609: ......