Mounting / changing the probe
61211-1270302
MMZ M
6-9
6 Direction in which the articulating probe holder is inserted into the dovetail
guide.
7 Hexagon socket screw, width across flats: 2.5 mm
1
Move the ram to a position from which the articulating system can
be attached to the ram.
Use the joysticks on the control console to move the CMM in all
axes.
2
Switch the drives off.
3
Slide the articulating system up to the stop into the dovetail guide on
the ram.
4
The adapter must be flush with the ram.
5
Secure the articulating system by tightening the hexagon socket
screw to a torque of 2.2 Nm.
6
Reactivate the drives.
Fastening the VAST
Slide the probe onto the dovetail guide of the ram. The probe must al-
ways be secured with a screw.
1
2
3
4
5
6
(3)
9
8
7
1 Ram
2 Groove for pin
3 Dovetail guide on the ram
4 Pin
5 Probe
6 Cover of the adapter plate receptacle