Maintenance
61211-1270302
MMZ M
9-3
Little maintenance:
–
Performance of preventive maintenance actions on all components
according to manufacturer specifications.
–
Check and correction of the geometry without laser measuring ma-
chine.
–
Visual check of the data system.
–
Test of the drives and motors in all axes.
–
Calibration and recording of measuring results.
Maintenance intervals
The time intervals between two maintenance calls depend on the de-
gree of utilization. Generally, maintenance should be carried out at least
every 2000 operating hours.
NOTE
ZEISS recommends performing little and complete maintenance on an
alternating basis: little maintenance in the first year and complete main-
tenance in the second year.
Components of maintenance work
Maintenance work includes inspection of the following:
Area
Examples
Safety devices (standard
equipment)
For example drive monitoring; collision
protection in the series measurement
mode
Measuring system
For example functioning, precision
Control
For example operator's controls; control
console
Drive unit
Motor and gear unit
Sensors
Probing system
NOTE
When concluding a maintenance agreement, you also must specify any
optional equipment to be covered.
Содержание MMZ M
Страница 18: ...Warranty 1 6 61211 1270302 MMZ M ...
Страница 65: ...4 61211 1270302 MMZ M 4 1 Technical data This chapter contains CMM 4 2 Technical data ...
Страница 74: ...Installation 5 6 61211 1270302 MMZ M ...
Страница 92: ...Start up checklist 6 18 61211 1270302 MMZ M ...
Страница 118: ...Terminating the measuring operation 7 26 61211 1270302 MMZ M ...
Страница 146: ...2Glossary 2 61211 1270302 MMZ M ...
Страница 152: ...6Alphabetic index 6 61211 1270302 MMZ M ...
Страница 153: ......
Страница 154: ...MMZ M Operating Instructions 2021 10 05 61211 1270302 ...