Care
61211-1270302
MMZ M
9-11
Control console
The control console can be wiped with a damp cloth. A mild cleaning
agent can be used here.
No humidity is allowed inside of the control console. For cleaning, use
only a damp cloth cloth. The cloth may not be wet.
Housing covers
Vacuum the housing covers and clean them with a damp cloth.
NOTE
The covers must not be removed.
Inspection measures
Checking the stylus system
Regular inspection of the stylus systems is required to ensure correct
workpiece measurement.
1
Check the styli at regular intervals.
2
Remove any particles or grease film from stylus tips.
3
Replace any damaged styli.
Cleaning or replacing the filter mats of the MCC 800
controller cabinet
The interior of the controller cabinet is cooled by a fan attached to the
rear side of the controller cabinet. The filter mats must be cleaned at
least every three months.
Verschmutztes
LUFTFILTER
austauschen
oder reinigen
CLEAN OR REPLACE
SOILED
AIR FILTER
2
3
1
MCC 800 controller cabinet; rear
1 Air filter
Содержание MMZ M
Страница 18: ...Warranty 1 6 61211 1270302 MMZ M ...
Страница 65: ...4 61211 1270302 MMZ M 4 1 Technical data This chapter contains CMM 4 2 Technical data ...
Страница 74: ...Installation 5 6 61211 1270302 MMZ M ...
Страница 92: ...Start up checklist 6 18 61211 1270302 MMZ M ...
Страница 118: ...Terminating the measuring operation 7 26 61211 1270302 MMZ M ...
Страница 146: ...2Glossary 2 61211 1270302 MMZ M ...
Страница 152: ...6Alphabetic index 6 61211 1270302 MMZ M ...
Страница 153: ......
Страница 154: ...MMZ M Operating Instructions 2021 10 05 61211 1270302 ...