CMM
61211-1270302
MMZ M
4-3
Fuse
230-240 V
C 16 A
100-125 V
25 A
Amount of heat generated
CMM
10,800 kJ/h
Data station and peripheral devices
3800 kJ/h
Compressed air
Certain requirements have to be met before connecting the compressed
air in order not to impair the function of the CMM.
Requirements
Pressure
6-10 bar
Air quality according to ISO 8573-1:2010
[6:4:4]
–
Air purity with regard to particle concentration ac-
cording to class 6
–
Air purity with regard to pressure dew point ac-
cording to class 4
–
Air purity with regard to oil concentration accord-
ing to class 4
Air consumption
Weight counterbalancing of the ram (only
for Z = 2000 mm)
80 Nl/min at 1 bar
Air consumption for RDS
8 Nl/min at 1 bar
Travel speed
Creep speed:
0 to 5 mm/s
Set-up mode: (MAN operating mode)
0 to 90 mm/s
Series measurement mode: (AUTO operating mode)
Axis
max. 300 mm/s
Vector (3D)
max. 520 mm/s
Содержание MMZ M
Страница 18: ...Warranty 1 6 61211 1270302 MMZ M ...
Страница 65: ...4 61211 1270302 MMZ M 4 1 Technical data This chapter contains CMM 4 2 Technical data ...
Страница 74: ...Installation 5 6 61211 1270302 MMZ M ...
Страница 92: ...Start up checklist 6 18 61211 1270302 MMZ M ...
Страница 118: ...Terminating the measuring operation 7 26 61211 1270302 MMZ M ...
Страница 146: ...2Glossary 2 61211 1270302 MMZ M ...
Страница 152: ...6Alphabetic index 6 61211 1270302 MMZ M ...
Страница 153: ......