CMM
4-4
61211-1270302
MMZ M
Environmental conditions
The following conditions must be met to ensure proper operation of the
coordinate measuring machine.
Operational readiness
Ambient temperature
10-35°C
Measuring operation
Relative humidity
40% - 70% (without condensation)
Room temperature
18-22°C
100 - 125 V: max. 3000 m
230 - 240 V: max. 2000 m
NOTE
Please see the separate specifications brochure for information about
permissible temperature variations.
Reference temperature
Temperature at which the promised measuring accuracy is guaranteed
when the conditions are met.
The following conditions must be met:
–
Temperature for measuring operation must be constant for at least
48 hours.
–
Temperature variations must not exceed certain values. Limit values:
2 K/d and 1 K/h.
Example of a reference temperature for measurement of 20°C:
Temperature
Change
Status
Action
Per day
18.1-20.0°C
1.9 K
OK
20.0-22.0°C
2.0 K
OK
17.9-20.0°C
2.1 K
Not OK
Wait for 48 hours until the new
temperature set is constant.
Per hour
20.9-20.0°C
0.9 K
OK
20.0-21.0°C
1.0 K
OK
18.9-20.0°C
1.1 K
Not OK
Wait for 48 hours until the new
temperature set is constant.
Содержание MMZ M
Страница 18: ...Warranty 1 6 61211 1270302 MMZ M ...
Страница 65: ...4 61211 1270302 MMZ M 4 1 Technical data This chapter contains CMM 4 2 Technical data ...
Страница 74: ...Installation 5 6 61211 1270302 MMZ M ...
Страница 92: ...Start up checklist 6 18 61211 1270302 MMZ M ...
Страница 118: ...Terminating the measuring operation 7 26 61211 1270302 MMZ M ...
Страница 146: ...2Glossary 2 61211 1270302 MMZ M ...
Страница 152: ...6Alphabetic index 6 61211 1270302 MMZ M ...
Страница 153: ......
Страница 154: ...MMZ M Operating Instructions 2021 10 05 61211 1270302 ...