Evaluation of the measuring results
7-24
61211-1270302
MMZ M
Source of error
Cause
Workpiece condition
–
The workpiece shows signs of deformation and roughness.
–
The workpiece is soiled.
–
The workpiece is covered with oil, for example drops of oil in the
small bores.
–
The workpiece material is too soft.
Magnetic field
–
A malfunction was caused by a magnetic field.
Possible causes of a magnetic field: for example workpiece, clamp-
ing tool, stylus system extensions.
Qualification
–
Different axis clamping set than for qualification.
–
Different measuring force set than for qualification (depending on
the measuring method, with measuring probes)
Probing
–
Probing has not been performed perpendicular to the probing sur-
face.
Содержание MMZ M
Страница 18: ...Warranty 1 6 61211 1270302 MMZ M ...
Страница 65: ...4 61211 1270302 MMZ M 4 1 Technical data This chapter contains CMM 4 2 Technical data ...
Страница 74: ...Installation 5 6 61211 1270302 MMZ M ...
Страница 92: ...Start up checklist 6 18 61211 1270302 MMZ M ...
Страница 118: ...Terminating the measuring operation 7 26 61211 1270302 MMZ M ...
Страница 146: ...2Glossary 2 61211 1270302 MMZ M ...
Страница 152: ...6Alphabetic index 6 61211 1270302 MMZ M ...
Страница 153: ......
Страница 154: ...MMZ M Operating Instructions 2021 10 05 61211 1270302 ...