Preparation for measuring operation
7-14
61211-1270302
MMZ M
4
Probe the reference sphere in the Z axis.
The reference sphere is measured, determining its position.
A Correct
B Incorrect
1 Stylus shaft
2 Probing direction
3 Reference sphere; probing on the pole.
5
Remove the master stylus.
All styli can now be qualified.
Qualification procedure
After the reference measurement has been carried out, the stylus sys-
tems can be qualified.
NOTE
During qualification, the temperature effect should be taken into ac-
count.
Prior to qualification, carry out temperature compensation.