Preparation for measuring operation
61211-1270302
MMZ M
7-9
Preparation for measuring operation
Prerequisites for a precise measuring run
The following preparations have to be made to ensure maximum accu-
racy when measuring with the CMM.
The CMM must be adapted to room temperature prior to the mea-
suring run. Recommendation: The CMM should always remain on.
1
Switch the CMM on at least
30 minutes
before starting the measur-
ing run.
The heat of the hand may influence the properties of probe holder
and probe. The components should therefore be installed before
starting the measurement.
2
Mount the probe to the CMM at least 15 minutes before starting the
measuring run.
3
Always carry out a reference measurement after the reference point
travel and qualify each stylus used for measurement.
4
Carry out temperature compensation.
➤
Homing run
Home position
The home position corresponds to the origin of the machine coordinate
system. It is located in the upper left corner of the measuring volume.
The home position must be determined by a homing run
prior
to the
measuring run.
+X
-Z
-Y
2
3
1
4
Содержание MMZ M
Страница 18: ...Warranty 1 6 61211 1270302 MMZ M ...
Страница 65: ...4 61211 1270302 MMZ M 4 1 Technical data This chapter contains CMM 4 2 Technical data ...
Страница 74: ...Installation 5 6 61211 1270302 MMZ M ...
Страница 92: ...Start up checklist 6 18 61211 1270302 MMZ M ...
Страница 118: ...Terminating the measuring operation 7 26 61211 1270302 MMZ M ...
Страница 146: ...2Glossary 2 61211 1270302 MMZ M ...
Страница 152: ...6Alphabetic index 6 61211 1270302 MMZ M ...
Страница 153: ......
Страница 154: ...MMZ M Operating Instructions 2021 10 05 61211 1270302 ...