Care
61211-1270302
MMZ M
9-5
Care
What you should know!
Although the CMM is robust and easy to care for, a certain degree of
care is still required. The sensitive measuring system of the CMM re-
quires a certain degree of cleanliness. Even the smallest dust particles on
the stylus tip or the reference sphere may cause inaccurate measure-
ments. Furthermore, the measuring plate requires daily care.
The measures for care also include the regular cleaning and checking of
certain components.
Cleaning agent:
Medium
Purpose
Order number
Mild cleaning agent
Contact 60
To pre-clean the contacts on
adapter plates and the adapter
plate receptacle of the probe car-
rier
NOTICE! Do not use for ball
pairs in the adapter plate re-
ceptacle.
000000-0118-669
Contact WL
For secondary cleaning of the
contacts on adapter plates and
the adapter plate receptacle of
the probe carrier
NOTICE! Do not use for ball
pairs in the adapter plate re-
ceptacle.
000000-0118-666 (75 ml)
000000-0119-013 (200 ml)
Lint-free cloth, for example
made of linen
Vacuum cleaner
Shellsol
Grease dissolving agent to clean
the drive surfaces and the mea-
suring plate
Corrosion preventing agent
Rivolta T.R.S. Plus
(pump
spray 300 ml)
To protect and clean the measur-
ing plate
000000-0518-313
Solvent
To remove aluminum deposits
from the stylus tip
Содержание MMZ M
Страница 18: ...Warranty 1 6 61211 1270302 MMZ M ...
Страница 65: ...4 61211 1270302 MMZ M 4 1 Technical data This chapter contains CMM 4 2 Technical data ...
Страница 74: ...Installation 5 6 61211 1270302 MMZ M ...
Страница 92: ...Start up checklist 6 18 61211 1270302 MMZ M ...
Страница 118: ...Terminating the measuring operation 7 26 61211 1270302 MMZ M ...
Страница 146: ...2Glossary 2 61211 1270302 MMZ M ...
Страница 152: ...6Alphabetic index 6 61211 1270302 MMZ M ...
Страница 153: ......
Страница 154: ...MMZ M Operating Instructions 2021 10 05 61211 1270302 ...