Malfunctions during measuring operations
8-4
61211-1270302
MMZ M
Malfunction
Cause
Action
Check
The default value set
in the measuring soft-
ware is too low, for
example 50 %.
–
Set the default speed value in
the measuring software to
100 %.
The SLOW function is
active.
–
Deactivate the SLOW function
via the control console.
Display on the control
console
The «MAN» operating
mode is set during an
automatic run.
Set the «AUTO» operating mode:
–
Turn the
key-operated switch
to
the «AUTO» position.
Stylus system
change
No holder has been
assigned to the stylus
system.
–
Assign a holder to the stylus
system.
The position of the
changer rack or
holder has changed.
–
Reassign the holder.
Collision: missing in-
termediate positions.
–
Reassign the holder and set in-
termediate positions.
Two stylus systems
have been assigned
to one holder.
–
Reassign the holder.