Preparation for measuring operation
61211-1270302
MMZ M
7-13
–
The limit values for assembly of the stylus system have been ob-
served.
–
The reference sphere, the stylus tips, and the adapter plate are clean
and not damaged.
–
Stylus system components have not been damaged.
Fastening the reference sphere holder
Place the reference sphere holder on the measuring plate and screw it
into a threaded hole in the measuring plate (max. torque 35 Nm).
NOTE
Select the position of the reference sphere holder so that the reference
sphere can be probed with all styli.
If it is not possible to qualify certain styli, you may change the orienta-
tion of the reference sphere. Example:
1 Vertical orientation of the reference sphere
Performing a reference measurement
The position of the reference sphere must be determined before the sty-
lus vector can be determined. Use the master stylus for this purpose.
Proceed as follows:
1
Insert the master stylus.
2
Open the menu for the reference measurement. Please see the user
guide for the measuring software.
3
Select stylus 1 on the control console.
Содержание MMZ M
Страница 18: ...Warranty 1 6 61211 1270302 MMZ M ...
Страница 65: ...4 61211 1270302 MMZ M 4 1 Technical data This chapter contains CMM 4 2 Technical data ...
Страница 74: ...Installation 5 6 61211 1270302 MMZ M ...
Страница 92: ...Start up checklist 6 18 61211 1270302 MMZ M ...
Страница 118: ...Terminating the measuring operation 7 26 61211 1270302 MMZ M ...
Страница 146: ...2Glossary 2 61211 1270302 MMZ M ...
Страница 152: ...6Alphabetic index 6 61211 1270302 MMZ M ...
Страница 153: ......
Страница 154: ...MMZ M Operating Instructions 2021 10 05 61211 1270302 ...