Safety on the CMM
2-18
61211-1270302
MMZ M
The
safe speed
option is activated if the light barrier is interrupted. After
that, the CMM travels at the same speed as in the MAN operating
mode. This means the execution of the measurement plan is continued
at reduced speed.
Returning to normal
speed
As soon as the CMM is outside the danger area, you can switch back to
normal speed using the switch on the signal tower. Then the CMM
moves at the speed according to the values defined in the measurement
plan.
Setting up the light barrier
To ensure that the light barrier serves as a safety device, it must be in-
stalled correctly. This includes the correct configuration and alignment.
In addition, the individual columns must be screwed to the floor to en-
sure that their position cannot be changed. When installing the light
barriers, make sure that the distance between the light beam and the
danger zone is at least 900 mm.
Position of control elements
All control elements must be located outside the blocked danger zone.
This applies to the control elements on the controller cabinet, control
console, and signal tower for the light barrier. Also ensure that the con-
trol elements cannot be reached from inside the blocked danger zone.
Ensure that users can view the entire danger zone from the location of
the signal tower for the light barrier.
Содержание MMZ M
Страница 18: ...Warranty 1 6 61211 1270302 MMZ M ...
Страница 65: ...4 61211 1270302 MMZ M 4 1 Technical data This chapter contains CMM 4 2 Technical data ...
Страница 74: ...Installation 5 6 61211 1270302 MMZ M ...
Страница 92: ...Start up checklist 6 18 61211 1270302 MMZ M ...
Страница 118: ...Terminating the measuring operation 7 26 61211 1270302 MMZ M ...
Страница 146: ...2Glossary 2 61211 1270302 MMZ M ...
Страница 152: ...6Alphabetic index 6 61211 1270302 MMZ M ...
Страница 153: ......
Страница 154: ...MMZ M Operating Instructions 2021 10 05 61211 1270302 ...