Safe operation of the CMM
61211-1270302
MMZ M
2-11
Home position
After start-up, travel movements may only be carried out to a limited ex-
tent using the joysticks located on the control console. Cause: No end
positions of the axes have been defined yet. Possible consequences:
Danger to personnel and damage to the CMM.
CAUTION
Risk of injury if the travel movement exceeds the end posi-
tions.
Crushing and severing body parts.
Damage to the CMM.
ü
Before performing unlimited travel movements, the end positions
must have been defined.
After start-up, first perform a homing run.
After this, travel movements can be performed without restrictions us-
ing the joysticks.
Suitable clothing
Certain work on the CMM presents a risk of injury caused by unsuitable
clothing. This applies particularly to the loading of the measuring plate
and the workpiece set-up on the measuring plate.
CAUTION
Risk of injury caused by heavy, sharp-edged workpieces.
Tearing of fingers and crushing of feet.
Wear safety shoes.
Wear safety gloves.
General precautions
Delimiting the work area
The work area around the CMM must be delimited by means of a light
barrier.
Do not modify the CMM
If the delivered version of the CMM is modified, safe operation of the
CMM is no longer guaranteed.
Содержание MMZ M
Страница 18: ...Warranty 1 6 61211 1270302 MMZ M ...
Страница 65: ...4 61211 1270302 MMZ M 4 1 Technical data This chapter contains CMM 4 2 Technical data ...
Страница 74: ...Installation 5 6 61211 1270302 MMZ M ...
Страница 92: ...Start up checklist 6 18 61211 1270302 MMZ M ...
Страница 118: ...Terminating the measuring operation 7 26 61211 1270302 MMZ M ...
Страница 146: ...2Glossary 2 61211 1270302 MMZ M ...
Страница 152: ...6Alphabetic index 6 61211 1270302 MMZ M ...
Страница 153: ......
Страница 154: ...MMZ M Operating Instructions 2021 10 05 61211 1270302 ...