Malfunctions during measuring operations
61211-1270302
MMZ M
8-3
Malfunctions during measuring operations
The following table helps you identify and eliminate malfunctions that
occur during measuring operations.
NOTE
A malfunction can have different causes. One or more measures may be
necessary to eliminate a certain cause. If two measures are specified,
this does not mean that you have to take both of them to fix the prob-
lem.
Malfunction
Cause
Action
Check
No travel
movement pos-
sible.
The
EMERGENCY
STOP
button on the
control console or
the controller cabinet
is locked.
Unlock the
EMERGENCY STOP
button:
–
Turn the button slightly.
The button pops out
a few millimeters.
Collision, for example
with the workpiece.
–
Eliminate the cause of the colli-
sion.
–
Reactivate the drives.
–
Reinitialize the CMM if neces-
sary.
«AUTO» operating
mode set or
key-op-
erated switch
in «0»
position.
–
Turn the
key-operated switch
to
«MAN».
A CNC function has
been activated and is
not yet complete,
e.g. a semi-automatic
measurement run.
–
Wait until the measurement run
is finished.
The CMM is in the
end position.
See
The pressure is too
low.
–
Increase the pressure in the
supply line.
Pressure gauges
The filter is clogged.
–
Change the filter.
Travel move-
ment too slow.
The set travel speed is
too low.
Increase the travel speed:
–
Turn the control knob on the
control console clockwise.
Содержание MMZ M
Страница 18: ...Warranty 1 6 61211 1270302 MMZ M ...
Страница 65: ...4 61211 1270302 MMZ M 4 1 Technical data This chapter contains CMM 4 2 Technical data ...
Страница 74: ...Installation 5 6 61211 1270302 MMZ M ...
Страница 92: ...Start up checklist 6 18 61211 1270302 MMZ M ...
Страница 118: ...Terminating the measuring operation 7 26 61211 1270302 MMZ M ...
Страница 146: ...2Glossary 2 61211 1270302 MMZ M ...
Страница 152: ...6Alphabetic index 6 61211 1270302 MMZ M ...
Страница 153: ......
Страница 154: ...MMZ M Operating Instructions 2021 10 05 61211 1270302 ...