Special measures
61211-1270302
MMZ M
8-5
Special measures
After a collision
If the ram or the probe collides with an obstacle, the guideway and
other components of the CMM may become damaged. In this case, per-
fect measuring operation is no longer guaranteed.
–
If measurements are not possible after a collision, call a ZEISS service
engineer or our support.
Notes
NOTE
After a collision, the stylus system has to be requalified and the work-
piece position (w position) has to be determined again.
NOTE
If measurements are not possible after a collision, call a ZEISS service en-
gineer or our support.
NOTE
After a collision, the stoplight switches to red in the stoplight window.
You have to switch back the stoplight to green before continuing mea-
surement. Please refer to the operating instructions for the measuring
software.
Collision between the probe and stylus
When the probe or stylus has collided with an obstacle and the drives
have been switched off, proceed as follows:
1
Eliminate the cause of the collision.
2
Activate the collision protection.
start
active
bypass
Turn the key-operated switch to «start» and hold until the corre-
sponding indicator lamp lights up.
3
Reactivate the drives.
4
Check whether the stylus needs to be requalified.
NOTE
If the drives cannot be switched on, the CMM must be reinitialized.
Please refer to the operating instructions for the measuring software.
Содержание MMZ M
Страница 18: ...Warranty 1 6 61211 1270302 MMZ M ...
Страница 65: ...4 61211 1270302 MMZ M 4 1 Technical data This chapter contains CMM 4 2 Technical data ...
Страница 74: ...Installation 5 6 61211 1270302 MMZ M ...
Страница 92: ...Start up checklist 6 18 61211 1270302 MMZ M ...
Страница 118: ...Terminating the measuring operation 7 26 61211 1270302 MMZ M ...
Страница 146: ...2Glossary 2 61211 1270302 MMZ M ...
Страница 152: ...6Alphabetic index 6 61211 1270302 MMZ M ...
Страница 153: ......
Страница 154: ...MMZ M Operating Instructions 2021 10 05 61211 1270302 ...