4-30
IM AQ1300-01EN
Method for saving continuance test result files
You can save each test item of a continuous test to separate files or save multiple test items to a
single file.
• Single: Each test item is saved to a separate file. This is the conventional method.
• Bundle: Selected test items are saved to a single file.
Note
• Files cannot be loaded into the setup software when the firmware version is older than R1.08.01.001,
regardless of whether Single or Bundle is selected.
• If a continuous test is stopped in the middle of execution (as a result of an error or pressing of the STOP
key), the test results up to that point are saved.
Measurement Screen
Measurement START Condition
Request ARP/NDP
For ARP (IPv4) and NDP (IPv6) destination MAC addresses, select whether or not to use an ARP
or NDP request to acquire the MAC address when the START key is pressed. This setting is valid
when the test layer is L3-IPv4 or L3-IPv6.
• Selected: An ARP/NDP request is sent.
• Cleared: An ARP/NDP request is not sent.
Not start the measurement when Link is down
| Select whether or not to start measurement when the START key is pressed and a linkdown is
detected.
• Selected: Measurement is not started when a linkdown is detected.
• Cleared: Measurement starts even when a linkdown is detected.
Measurement STOP Condition
When Link Down is detected
Select whether or not to stop measurement when a linkdown is detected during measurement.
• Selected: Measurement is stopped when a linkdown is detected.
• Cleared: Measurement is not stopped even when a linkdown is detected.
When L2 Error is detected
Select whether or not to stop measurement when an L2 error is detected during measurement.
• Selected: Measurement is stopped when an L2 error is detected.
• Cleared: Measurement is not stopped even when an L2 error is detected.
When L3 Error is detected
Select whether or not to stop measurement when an L3 error is detected during measurement.
• Selected: Measurement is stopped when an L3 error is detected.
• Cleared: Measurement is not stopped even when an L3 error is detected.
When Test Fail is detected
Select whether or not to stop continuance test when a test fail judgment is detected during
continuance test. This feature is supported in firmware version (FW Ver.) R1.08.01.001 and later.
• Selected: Continuance test is stopped when a test fail judgment is detected.
• Cleared: Continuance test is not stopped even when a test fail judgment is detected.
4.11 Configuring Options (Auto)