7-6
IM AQ1300-01EN
Explanation
When Measurement and Transmission Control Are Synchronized
Start and stop measurement this way when you want to perform measurement and transmission at the
same time.
START key: Press this key to start measurement and transmission.
STOP key: Press this key to stop transmission and finish measurement.
When the transmission end conditions are met, transmission stops automatically, and measurement
finishes.
Note
When Auto(Remote) has been selected in the Test menu, measurement and transmission control are
synchronized regardless of the measurement and transmission control synchronization setting.
When Measurement and Transmission Control Are Not
Synchronized
Start and stop measurement this way when you want to perform measurement and transmission
separately.
START key:
Press this key to start measurement.
Start Transmit soft key:
Press this soft key to start transmission.
Stop Transmit soft key:
Press this soft key to stop transmission.
STOP key:
Press this key to stop measurement.
When the transmission end conditions are met, transmission stops automatically.
Note
If you press STOP without pressing the Stop Transmit soft key first, transmission and measurement are both
stopped.
When Auto or Auto(Remote) Has Been Selected, Execution Type
Is Set to Continue, and Continuance Confirmation Is Enabled
When Auto or Auto(Remote) has been selected in the Test menu, the test item Execution Type
setting is set to Continue, and Continuance Confirmation is set to ON, a confirmation screen appears
whenever a test item ends during measurement.
Performing Pass/Fail Judgment When Auto or Auto(Remote) Has
Been Selected
This indication appears when the test type is Auto or Auto(Remote) and you have chosen to perform
pass/fail judgment.
• Pass: The results meet the requirements.
• Fail: The results do not meet the requirements.
7.1 Starting and Stopping Measurement