6-53
IM AQ1300-01EN
Manual T
est Settings (Manual)
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
App
Index
Explanation
Advance Setup Screen
Synchronize Measurement and Transmission controls
Select whether or not to start transmission when the START key is pressed.
• Selected: Transmission is started when the START key is pressed. Also, measurement is
stopped when transmission finishes.
• Cleared: Transmission is not started when the START key is pressed.
Wait time before starting transmission
Set the interval between when the START key is pressed and when transmission is started and
between when transmission finishes and when measurement is stopped. This setting is valid when
the “Synchronize Measurement and Transmission controls” check box is selected.
Range: 1 to 10 s
Remove Test Tag from Test Frame
When the test mode is Traffic Test or QoS Test, you can send test frames without test tags.
• Selected: The test tags are removed from test frames.
• Cleared: The test tags are not removed from test frames.
Note
If you select the “Remove Test Tag from Test Frame” check box, the following operations may not work
properly.
• Payload error detection
• Sequence error measurement
• Latency measurement
• Frame ID classification in QoS measurement
Measurement Screen
Measurement START Condition
Request ARP/NDP
For ARP (IPv4) and NDP (IPv6) destination MAC addresses, select whether or not to use an ARP
or NDP request to acquire the MAC address when the START key is pressed. This setting is valid
when the test layer is L3-IPv4 or L3-IPv6.
• Selected: An ARP/NDP request is sent.
• Cleared: An ARP/NDP request is not sent.
Not start the measurement when Link is down
Select whether or not to start measurement when the START key is pressed and a linkdown is
detected.
• Selected: Measurement is not started when a linkdown is detected.
• Cleared: Measurement starts even when a linkdown is detected.
Measurement STOP Condition
When Link Down is detected
Select whether or not to stop measurement when a linkdown is detected during measurement.
• Selected: Measurement is stopped when a linkdown is detected.
• Cleared: Measurement is not stopped even when a linkdown is detected.
When L2 Error is detected
Select whether or not to stop measurement when an L2 error is detected during measurement.
• Selected: Measurement is stopped when an L2 error is detected.
• Cleared: Measurement is not stopped even when an L2 error is detected.