9-12
IM AQ1300-01EN
9.6 Displaying
Back to Back Test Results
Procedure
Back to Back Test Results Screen
Follow the procedure in section 9.1 to start and stop measurement.
Press the
Back to Back
soft key to display the following screen.
Back to back test results graph
Back to back test results table
Note
• You can display this screen even during measurement.
• You can also use the left and right arrow keys to switch between measurement screens. For details, see
section 9.8.
• Press ESC to return to the Test Result Display screen (see section 9.2).
Explanation
Back to Back Test Results Graph
The measured results are displayed on a graph.
• Horizontal axis: Test frame length
• Vertical axis:
Number of burst frames
Back to Back Test Results Table
The following items are displayed.
• No: This is the test frame length number. If pass/fail judgment is enabled, the pass/fail
judgment results are indicated with colors for each frame length.
Green: Pass
Red: Fail
• Length: Test frame length
• NumOfFrames: The measured results are displayed as the number of burst frames.
Note
• Use the up and down arrow keys to switch the displayed page. If more than seven test frame lengths
are enabled, the first seven frame lengths are displayed on the first page, and the remaining test frame
lengths are displayed on the second page.
• During measurement or for any test frame lengths that have not yet gone through measurement, the
result columns display “----.”