2-2
IM AQ1300-01EN
Title Bar
Simple Optical Power Monitor
When the interface is XFP or SFP, the AQ1300/AQ1301 can monitor the received optical power
level and indicate it using three levels.
Indication Colors
Description
Black and black When the RJ-45 interface is being used or when the XFP or SFP interface is
being used and the interface module is not installed
Red and black
When the level of received light is too low
Green and green When the level of received light is appropriate
Red and red
When the level of received light is too high
If you are using optical modules that are not recommended on an AQ1300/AQ1301 whose firmware
version is R1.05.01.001 or later, black and black is displayed.
Current Time
The current time is displayed (YYYY/MM/DD hh:mm:ss).
Selected Setup File
When Auto or Auto(Remote) has been selected in the Test menu, the comment or file name that has
been specified by the setup file selection is displayed.
When Manual has been selected in the Test menu, the file name is displayed.
Displayed characters: Up to 30
Note
When Auto(Remote) has been selected in the Test menu and the screen is sent from the master to the slave,
“Send Screen Image” appears on the slave.
Link Status Indication
The link status is indicated.
Indication Color
Description
Blue
Linkdown
Green
Linkup
Power Supply Status Indication
The power supply status is indicated.
Indication Color
Description
—
The AC adapter is being used.
Green
The battery is sufficiently full.
Yellow
The battery is about half full.
Red
The battery is almost empty.
Test Information
The test type, mode, and item execution status are indicated.
Information Type
Indication
Description
Test type
Auto
Auto mode
Auto (RMT)
Auto(Remote) mode
Manual
Manual mode
Test mode
Traffic
Traffic mode
Loopback
Loopback mode
QoS
QoS test mode
PING
PING test mode
BERT
BERT mode
Test item execution
status
The test type is Auto or Auto(Remote), and Execution Type is set to
Single.
The test type is Auto or Auto(Remote), and Execution Type is set to
Continue.
1 to 8
The number of the test item that is currently being executed or is
currently selected. The number blinks when the test item is being
executed.
2.1 Measurement Screens (Common Items)