Rx Setting Screen
Oversize Threshold
Set the threshold for determining oversize frames.
Range: 65 to 10,000 bytes
Detect Payload Error
Select whether or not to detect payload errors.
• Selected: Payload errors are detected.
• Cleared: Payload errors are not detected.
Note
If you select the “Remove Test Tag from Test Frame” check box in the Option (Manual) Advance setup, the
AQ1300/AQ1301 will be unable to detect payload errors.
•
Offset
Set the payload error offset.
• Match to Tx Setting: The offset is based on the source setting.
• Manual: You must set the offset manually.
•
Offset Byte Setting
Specify the offset byte for when Offset is set to Manual.
Range: Byte 0 to 255
Target Frame Setting
Set which frames to take statistics of.
• Test Frame Only: Statistics are only taken for test frames.
• All Frames:
Statistics are taken for all frames.
Note
When you are using an Rx base filter, statistics are taken of the frames that pass through the filter.
Rx Base Filter Setting
•
Frame Structure
The specified frame structure is displayed.
Detail
Press to configure the Rx frame settings in the Frame Setting screen. For details, see section 6.5.
•
Filter 1 and 2
Select whether or not to use Rx base filters 1 and 2.
• Selected: The Rx base filter is used.
• Cleared: The Rx base filter is not used.
The selected Rx base filter type is displayed.
Detail
Press to configure the Rx base filter settings in the Rx Base Filter Setting screen. For details, see
section 6.5.