WARNING
Inaccurate measurement risk. Use only Masimo rainbow SET
sensors and accessories on Masimo-equipped monitors.
WARNING
Inaccurate measurement risk. Use only Nellcor sensors and
accessories on Nellcor-equipped monitors.
WARNING
High-intensity lights (including pulsating strobe lights) directed
on the pulse co-oximeter can prevent it from obtaining readings.
WARNING
The pulsations from intra-aortic balloon support can increase
the pulse rate displayed on the monitor. Verify the patient's pulse rate
against the ECG heart rate.
WARNING
Inaccurate measurement risk. Venous congestion might cause
a low reading of arterial oxygen saturation. Place the sensor on a hand at
heart level to ensure proper venous outflow from a monitored site.
WARNING
Inaccurate measurement risk. The pulse co-oximeter can be
used during defibrillation, but the reading may be inaccurate for up to 20
seconds.
WARNING
Patient injury risk. Do not attempt to reprocess, recondition, or
modify any sensors or patient cables. Doing so might damage electrical
components.
WARNING
Pulse rate measurement might not detect certain arrhythmias
because it is based on the optical detection of a peripheral flow pulse. Do
not use the pulse oximeter as a replacement or substitute for ECG-based
arrhythmia analysis.
WARNING
Use the pulse co-oximeter as an early warning device. As you
observe a trend toward patient hypoxemia, use laboratory instruments to
analyze blood samples to better understand the patient's condition.
WARNING
Patient injury risk. If the sensor is applied with excessive
pressure for prolonged periods, a pressure injury can occur.
WARNING
Functional testers cannot be used to assess the accuracy of a
pulse oximeter monitor.
Directions for use
Patient monitoring 167
Содержание Connex IWS
Страница 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Страница 10: ...4 Introduction Welch Allyn Connex Devices...
Страница 14: ...8 Symbols Welch Allyn Connex Devices...
Страница 20: ...14 Screen elements Welch Allyn Connex Devices...
Страница 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Страница 72: ...66 Navigation Welch Allyn Connex Devices...
Страница 96: ...90 Profiles Welch Allyn Connex Devices...
Страница 106: ...100 Patient data management Welch Allyn Connex Devices...
Страница 120: ...114 Alarms Welch Allyn Connex Devices...
Страница 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Страница 212: ...206 Advanced settings Welch Allyn Connex Devices...
Страница 244: ...238 Specifications Welch Allyn Connex Devices...
Страница 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Страница 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Страница 266: ......