Set up etCO2
Follow these steps to configure etCO2 parameter settings and to set up etCO2
modifiers.
1. Touch the
Settings
tab.
2. Touch the
Setup
tab.
3. Touch the
etCO2
vertical tab.
4. Adjust the following settings as desired:
•
Sweep speed. Select the desired speed. You can set the default in Advanced
settings.
•
Waveform scale. Select the scale using the toggle button. You can set the
default in Advanced settings.
•
Pump standby time. Select the time using the toggle button.
5. Touch the
Home
tab.
The new settings take effect immediately.
6. To set up modifiers, if applicable, touch the
Patients
tab.
7. Touch the
Manual
tab.
8. Scroll through the list to find the etCO2 section, then enter or select the etCO2
modifier(s) as desired.
Note
If no modifiers are associated with this parameter, no etCO2
section will appear.
9. Touch
OK
.
The Home tab appears. In the Continuous Monitoring profile, the modifiers are saved
with the next set of measurements sent to the network.
Configure etCO2 and FiCO2 alarms
Follow these steps to set alarm limits for CO2 measurements.
Note
FiCO2 appears only if enabled in Advanced settings.
1. Verify that you are using the Continuous Monitoring profile.
Note
If “Allow profile change” is enabled in Advanced settings,
the device attempts to shift automatically to the Continuous
Monitoring profile when you connect the sensor cable.
2. Touch the
Alarms
tab.
3. Touch the
etCO2
vertical tab.
4. Ensure that the etCO2 and FiCO2 alarm limit controls are set to ON.
Note
If any parameter's alarm limit control is set to OFF, you
cannot adjust alarm limits on the Alarm tab, and no visual or
audio signals occur for that specific parameter.
5. Adjust the following settings as desired:
•
etCO2 limits. Enter upper and lower alarm limits for etCO2 using the up/down
arrow keys or the keypad next to the Limits control.
Directions for use
Patient monitoring 123
Содержание Connex IWS
Страница 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Страница 10: ...4 Introduction Welch Allyn Connex Devices...
Страница 14: ...8 Symbols Welch Allyn Connex Devices...
Страница 20: ...14 Screen elements Welch Allyn Connex Devices...
Страница 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Страница 72: ...66 Navigation Welch Allyn Connex Devices...
Страница 96: ...90 Profiles Welch Allyn Connex Devices...
Страница 106: ...100 Patient data management Welch Allyn Connex Devices...
Страница 120: ...114 Alarms Welch Allyn Connex Devices...
Страница 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Страница 212: ...206 Advanced settings Welch Allyn Connex Devices...
Страница 244: ...238 Specifications Welch Allyn Connex Devices...
Страница 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Страница 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Страница 266: ......