Symptom
Possible cause
Suggested action
The lamp brightness does not adjust
The platform handle controller PCBA is
defective
Call service.
The handle assembly is defective
Call service.
The handle becomes very hot to the
touch
The lamp has been on for an extended
period of time
Return the handle to the cradle.
Patient data management messages
Message
Possible cause
Suggested action
Maximum number of patient
records saved. Oldest record
overwritten.
The maximum number of patient records in the
monitor's memory has been exceeded
On the Review tab, delete old
records to prevent the alarm from
appearing when new records are
saved.
No data saved.
No patient data is available
Take or enter vital signs before
saving.
Patient ID required to save data.
The configuration requires a patient ID to save
data
Call for service to modify
advanced settings.
Clinician ID required to save data. The configuration requires a clinician ID to save
data
Call for service to modify
advanced settings.
Patient ID required to send data.
The configuration requires a patient ID to send
data
Add a patient ID.
Patient list is full. Delete some
patients to add more.
The maximum number of patients was exceeded Delete a patient from the list to
add a new patient.
Stop intervals to select new
patient.
The monitor is set to take interval readings
Stop intervals before changing
the patient.
No connection for send.
No connectivity is available to support sending
data manually or automatically sending data on
manual save
Call for service to check the
network connection or wireless
settings.
Unable to retrieve list.
The monitor is unable to retrieve a patient list
from the network
Call for service to check the
network connection or wireless
settings, or to verify that the
server is available.
Unable to identify clinician.
The clinician ID or password is incorrect
Confirm the clinician ID and
password (if applicable), and
retry.
Unable to identify patient.
Touch Clear to delete all data.
The patient ID does not match an ID in the
patient list or network
Re-enter the patient ID.
To delete all unsaved data, touch
Clear
.
Unable to identify clinician.
The clinician ID does not match an ID in the
network
Re-enter the clinician ID.
Directions for use
Troubleshooting 215
Содержание Connex IWS
Страница 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Страница 10: ...4 Introduction Welch Allyn Connex Devices...
Страница 14: ...8 Symbols Welch Allyn Connex Devices...
Страница 20: ...14 Screen elements Welch Allyn Connex Devices...
Страница 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Страница 72: ...66 Navigation Welch Allyn Connex Devices...
Страница 96: ...90 Profiles Welch Allyn Connex Devices...
Страница 106: ...100 Patient data management Welch Allyn Connex Devices...
Страница 120: ...114 Alarms Welch Allyn Connex Devices...
Страница 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Страница 212: ...206 Advanced settings Welch Allyn Connex Devices...
Страница 244: ...238 Specifications Welch Allyn Connex Devices...
Страница 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Страница 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Страница 266: ......