set to 0, 10, 25, 50, or 100 SatSeconds. If a desaturation event resolves on its own
within the preset time, the clock will automatically reset and the monitor will not alarm.
Note
The SatSeconds feature has a built-in safety protocol that sounds an alarm
whenever three SpO2 violations of any amount or duration occur within a
1-minute period.
SpO2 waveform view
The waveform view shows the SpO2 plethysmograph waveform. You can select the
default SpO2 waveform sweep speed in Advanced settings, but you can change the
sweep speed on the Setup tab.
For additional normalization and waveform information, refer to the manufacturer's
directions for use.
Set up SpO2
Follow these steps to configure SpO2 parameter settings and to set up modifiers.
1. Touch the
Settings
tab.
2. Touch the
Setup
tab.
3. Touch the
SpO2
vertical tab.
4. Select the desired Sweep speed.
5. Touch the
Home
tab.
The new setting takes effect immediately.
6. To set up modifiers, touch the
Patients
tab.
7. Touch the
Manual
tab.
8. Scroll through the list to find the SpO2 section, then enter or select the SpO2
modifier(s) as desired.
•
SpO2 measurement site. Select the measurement site from the list box.
•
O2 flow rate. Enter the flow rate using the keypad.
Note
All O2 modifier selections are cleared when the O2 flow
rate is changed to 0 and when the O2 method is
changed to none.
•
O2 concentration. Enter the concentration using the keypad.
•
O2 Method. Select the method from the list box.
Directions for use
Patient monitoring 165
Содержание Connex IWS
Страница 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Страница 10: ...4 Introduction Welch Allyn Connex Devices...
Страница 14: ...8 Symbols Welch Allyn Connex Devices...
Страница 20: ...14 Screen elements Welch Allyn Connex Devices...
Страница 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Страница 72: ...66 Navigation Welch Allyn Connex Devices...
Страница 96: ...90 Profiles Welch Allyn Connex Devices...
Страница 106: ...100 Patient data management Welch Allyn Connex Devices...
Страница 120: ...114 Alarms Welch Allyn Connex Devices...
Страница 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Страница 212: ...206 Advanced settings Welch Allyn Connex Devices...
Страница 244: ...238 Specifications Welch Allyn Connex Devices...
Страница 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Страница 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Страница 266: ......