Note
New batteries are only 30 percent charged. Therefore, connect the monitor
to AC power immediately after inserting a new battery.
WARNING
Risk of fire, explosion, and burns. Do not short-circuit, crush,
incinerate, or disassemble the battery pack. Never dispose of batteries in
refuse containers. Always recycle batteries according to local regulations.
Remove the wall system from the wall (Connex IWS)
For maintenance or service activities that require access to the back of the wall system,
follow these steps to remove the wall system from the wall.
1. Touch the
Settings
tab.
2. Touch the
Device
tab.
3. Touch
Power down
.
4. Remove all instrument heads, detach all accessible cords and cables, and unplug the
power cord from the outlet.
5. Remove the cover by loosening the captive retention screws.
6. If USB accessories are connected, loosen the two screws on the cable retention
clamp and remove it, then disconnect all USB cables.
7. If the wall system is configured with SpO2, disconnect the SpO2 cable and remove
it from the channel on the bottom of the wall system.
182 Maintenance and service
Welch Allyn Connex
®
Devices
Содержание Connex IWS
Страница 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Страница 10: ...4 Introduction Welch Allyn Connex Devices...
Страница 14: ...8 Symbols Welch Allyn Connex Devices...
Страница 20: ...14 Screen elements Welch Allyn Connex Devices...
Страница 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Страница 72: ...66 Navigation Welch Allyn Connex Devices...
Страница 96: ...90 Profiles Welch Allyn Connex Devices...
Страница 106: ...100 Patient data management Welch Allyn Connex Devices...
Страница 120: ...114 Alarms Welch Allyn Connex Devices...
Страница 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Страница 212: ...206 Advanced settings Welch Allyn Connex Devices...
Страница 244: ...238 Specifications Welch Allyn Connex Devices...
Страница 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Страница 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Страница 266: ......