Note
To switch to the Direct mode, touch
after the
Predictive mode measurement is acquired. The temperature
frame displays text indicating that the Direct mode is active.
The monitor sounds a tone to indicate the start of a Direct
measurement. Once you are in the Direct mode, continue to
separate the buttocks and hold the probe in place throughout
the measurement process.
Note
The monitor does not retain Direct mode temperatures in
memory. Therefore, it is important to note the temperature
before removing the probe from the measurement site and
then manually record it in the patient record.
6. Remove the probe after the temperature measurement is complete and firmly press
the eject button on the top of the probe to release the probe cover.
7. Return the probe to the probe well.
8. Wash your hands to reduce the risk of cross-contamination.
Braun ThermoScan
®
PRO thermometer and accessory dock
The thermometer and accessory dock enable you to transfer an ear temperature
measurement to the monitor. The dock also charges the thermometer battery.
Read the thermometer manufacturer's directions for use before attempting to configure,
use, troubleshoot, or maintain the thermometer.
WARNING
Liquids can damage electronics inside the thermometer.
Prevent liquids from spilling on the thermometer. If liquids are spilled on
the thermometer, dry off the thermometer with a clean cloth. Check for
proper operation and accuracy. If liquids possibly entered the thermometer,
remove the thermometer from use until it has been properly dried,
inspected, and tested by qualified service personnel.
WARNING
Many environmental variables, including patient physiology and
clinical application, can affect the accuracy and performance of the monitor.
Therefore, you must verify all vital signs information, especially
Temperature, before treating the patient. If there is any question about the
accuracy of a measurement, verify the measurement using another
clinically accepted method. If the thermometer configured with the device
is unavailable for any reason, use a different thermometer.
CAUTION
Probe covers are disposable, nonsterilized, and single-use. The
thermometer is also nonsterilized. Do not autoclave the thermometer and
probe covers. Ensure that probe covers are disposed of according to facility
requirements or local regulations.
CAUTION
The thermometer has no user-serviceable parts. If service is
required, call your nearest Welch Allyn Customer Service or Technical
Support facility.
Directions for use
Patient monitoring 159
Содержание Connex IWS
Страница 1: ...Welch Allyn Connex Devices Directions for use Software version 2 3X...
Страница 10: ...4 Introduction Welch Allyn Connex Devices...
Страница 14: ...8 Symbols Welch Allyn Connex Devices...
Страница 20: ...14 Screen elements Welch Allyn Connex Devices...
Страница 28: ...22 About warnings and cautions Welch Allyn Connex Devices...
Страница 72: ...66 Navigation Welch Allyn Connex Devices...
Страница 96: ...90 Profiles Welch Allyn Connex Devices...
Страница 106: ...100 Patient data management Welch Allyn Connex Devices...
Страница 120: ...114 Alarms Welch Allyn Connex Devices...
Страница 184: ...178 Patient monitoring Welch Allyn Connex Devices...
Страница 212: ...206 Advanced settings Welch Allyn Connex Devices...
Страница 244: ...238 Specifications Welch Allyn Connex Devices...
Страница 250: ...244 Standards and compliance Welch Allyn Connex Devices...
Страница 252: ...246 Guidance and manufacturer s declaration Welch Allyn Connex Devices...
Страница 266: ......